Currently Levelmeters display the He level only as a mapping from the measured fill height in % to the volume of the Vessel. On thge serverside a proper calibration using a lookup table is already implemented.
Sometihng similar is needed for displaying corrected He levels on the device itself.
The LUT:
- 11 entrys
- steps for 0% -100% in 10% steps
- this means ther will be 10 bins for interpolation
Interpolation:
y_LevelCal = y_1 + (x_LevelMeasured - x_1) * (y_2 - y_1) / (x_2 - x_1)
example lut:
| Level |
Vessel Vol |
| 0% |
10L |
| 10% |
20L |
| 20% |
30L |
| 30% |
60L |
| 40% |
90L |
| 50% |
120L |
| 60% |
150L |
| 70% |
180L |
| 80% |
190L |
| 90% |
200L |
| 100% |
210L |
- Sending the LUT from the Server will necessetate a new message code
- it will be sent some time during startup
- LUT will be saved in EEPROM
- LUT functionality can be enabled in Options (external pressure sensor option can be removed)